ZONAREz.one ultra and z.one ultra sp Ultrasound SystemsOperator Manual
z.one ultra and z.one ultra sp Systemsx z.one ultra and z.one ultra sp OPERATOR MANUALAcoustic Artifacts...
6 6-22 z.one ultra and z.one ultra sp OPERATOR MANUALC9-3Applications:1st & 2nd Trimester ObstetricsPediatric/Small Adult AbdomenPeripheral Vas
6z.one ultra and z.one ultra sp OPERATOR MANUAL 6-23L10-5Applications:ThyroidBreastScrotumPediatric HipsPeripheral VascularSuperficial AnatomyOcular
6 6-24 z.one ultra and z.one ultra sp OPERATOR MANUALWith the ZONARE Ultra ultrasound system, the L10-5, L14-5sp, P9-3ic, and P10-4 transducers are
6z.one ultra and z.one ultra sp OPERATOR MANUAL 6-25WARNING: To avoid injury to the patient, use only the Ocular Preset when performing imaging thr
This page intentionally blank
z.one ultra and z.one ultra sp OPERATOR MANUAL 7-177ImagingPatient Information FormThe Patient Information form is the record of the exam (patient ID
7 7-2 z.one ultra and z.one ultra sp OPERATOR MANUALSoftkeys The softkeys at the bottom of the Patient Information page have the following function
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-3To quickly find a patient name in the Worklist Selection Table, click the Query Patient softkey.
7 7-4 z.one ultra and z.one ultra sp OPERATOR MANUAL4. When the Patient Information form is complete, select Exit or press Freeze to save the form
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-5WARNING: To avoid injury to the patient, use only the Ocular Preset when performing imaging thro
z.one ultra and z.one ultra sp OPERATOR MANUAL xiz.one ultra and z.one ultra sp SystemsFiguresFigure 2-1. z.one ultra Convertible Ultrasound System ..
7 7-6 z.one ultra and z.one ultra sp OPERATOR MANUAL5. To save and exit:a. When the form is complete, select Exit or press Freeze to save the form
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-7NOTE: Choose a Preset that is the closest to your desired Preset parameters.4. Press the Menu bu
7 7-8 z.one ultra and z.one ultra sp OPERATOR MANUAL To Enable/Disable a Factory or User Preset in MenuNOTE: Factory Presets are identified by a
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-9Figure 7-5. Preset Customization Configuration Screen3. Select a transducer in the Transducer dr
7 7-10 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 7-6. Function Key Assignments Screen To Configure Function Keys1. Go to: 2. The Funct
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-114. Select the function you want to assign to the Function Key. Then click Apply.NOTE: There are
7 7-12 z.one ultra and z.one ultra sp OPERATOR MANUALImage Width (Live) (SmartCart Only)Image Width enables users to quickly narrow the lateral mar
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-133. For Stopwatch RESET/PHASE, see “Stopwatch RESET/PHASE Softkey” on page 7-76.NOTE: You can se
7 7-14 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 7-8. B-Mode Image Display FormatFigure 7-9. B-Mode Menu (Imaging Tab) Working with B-
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-15 To perform retrospective processing on a frozen image, see “Live and Frozen Images” on page 7-1
z.one ultra and z.one ultra sp Systemsxii z.one ultra and z.one ultra sp OPERATOR MANUALFigure 5-1. Scan Engine LCD Display...
7 7-16 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 7-12. Linear and Virtual Apex Array Icons and SoftkeysDual Mode Turning on Dual imagin
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-174. Optimize orientation on patient’s anatomy, user settings, and image quality for this second i
7 7-18 z.one ultra and z.one ultra sp OPERATOR MANUALNOTE: A transducer must be attached to the Scan Engine for the Optimize feature to work. The
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-19Optimize Button Operation To Activate Automatic Overall/DGC Gain Optimization To Activate Soun
7 7-20 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 7-14. Menu Access - Compounding Function To Activate Image Compounding (SmartCart/Sma
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-21Table 7-1. B-Mode Controls B ControlSmartCart/SmartCart sp Scan Engine✱ Indicates this control
7 7-22 z.one ultra and z.one ultra sp OPERATOR MANUALHarmonics SmartCartSmartCart spScan EngineSmartCart Press TH button to toggle On/Off tissue h
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-23ZoomSmartCart/SmartCart spScan Engine Acoustic Zoom (During live scanning): Press Zoom button.
7 7-24 z.one ultra and z.one ultra sp OPERATOR MANUALDynamic Range✱SmartCart/SmartCart sp Press the Menu button. Highlight Dyn Range with the Trac
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-25DGC ✱SmartCartSmartCart spScan EngineSmartCart Move a DGC slider to the right to increase gain,
1z.one ultra and z.one ultra sp OPERATOR MANUAL xiiiz.one ultra and z.one ultra sp SystemsFigure 7-21. PW Autotrace Icon - Doppler Data Field ...
7 7-26 z.one ultra and z.one ultra sp OPERATOR MANUALCompounding✱SmartCart/SmartCart spSmartCartNOTE: All ZONARE ultrasound factory Presets have C
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-27Dual: OFF/ONSmartCartSmartCart spScan EngineNOTE: The Dual imaging function allows for displayi
7 7-28 z.one ultra and z.one ultra sp OPERATOR MANUALSimul: (Dual)Scan Engine Press the softkey assigned Simul to toggle On/Off the Simultaneous D
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-29Motion Mode (M-Mode) ImagingMotion mode (M-Mode) is used to quantify the dynamic movement of tis
7 7-30 z.one ultra and z.one ultra sp OPERATOR MANUALa. Press M button. Press Menu button (SmartCart/SmartCart sp) or Menu Control (Scan Engine) to
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-31DepthSmartCart/SmartCart spScan Engine Press the Depth rocker button Up to decrease depth; pres
7 7-32 z.one ultra and z.one ultra sp OPERATOR MANUALDisp Format ✱ Press Menu button. Highlight Disp Format (display format) with Trackball. Move
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-33Dynamic Range✱ Press Menu button. Highlight Dyn Range with Trackball. Move Trackball left/righ
7 7-34 z.one ultra and z.one ultra sp OPERATOR MANUALColor and Power Doppler ImagingColor Doppler imaging provides qualitative directional informat
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-35Working with Color and Power Doppler Imaging To Access Color Doppler Imaging Mode1. Acquire the
z.one ultra and z.one ultra sp Systemsxiv z.one ultra and z.one ultra sp OPERATOR MANUALFigure 9-24. Auto-Dop Trace - Automeasure (Single Cycle)...
7 7-36 z.one ultra and z.one ultra sp OPERATOR MANUALDual Imaging ModeDual imaging mode enables the ultrasound system to display two side-by-side i
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-373. To toggle to (activate) the other Dual image (B):4. Optimize the orientation on the patient’s
7 7-38 z.one ultra and z.one ultra sp OPERATOR MANUALColor Doppler and Power Doppler ControlsTable 7-3. Color Doppler and Power Doppler Controls C
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-39Pwr DopplerScan Engine Press Pwr Dop softkey to toggle between Color Doppler and Power Doppler
7 7-40 z.one ultra and z.one ultra sp OPERATOR MANUALZoomThere are two types of Zoom. Acoustic Zoom requires that the image be “live” or in realtim
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-41Frequency ✱SmartCartSmartCart spScan Engine Press Frequency button to cycle through the transmi
7 7-42 z.one ultra and z.one ultra sp OPERATOR MANUALDual: OFF/ONSmartCartSmartCart spScan EngineSmartCart Press Dual button to toggle Dual imagin
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-43Pulsed Wave (PW) Doppler ImagingPulse Wave (PW) Doppler is used to assess blood flow velocity.Fi
7 7-44 z.one ultra and z.one ultra sp OPERATOR MANUALWorking with PW Doppler Imaging To Initiate PW Doppler Imaging1. Acquire the B-Mode image. Se
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-45PW Doppler ControlsTable 7-4. PW Doppler Controls PW Control SmartCart/SmartCart sp Scan Engin
1z.one ultra and z.one ultra sp OPERATOR MANUAL xvz.one ultra and z.one ultra sp SystemsFigure 12-6. Printer Setup Screen: Local/USB ...
7 7-46 z.one ultra and z.one ultra sp OPERATOR MANUALUpdate ✱Scan Engine Press Update softkey to toggle between live update of B-Mode image concu
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-47Tints SmartCart/SmartCart sp Press Menu button. Highlight Tints. Move Trackball left/right (or
7 7-48 z.one ultra and z.one ultra sp OPERATOR MANUALSweep Speed ✱SmartCart/SmartCart spNOTE: See “Sweep Speed” on page 10-7 for data on sweep s
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-49ZoomSmartCart/SmartCart spScan Engine Press Zoom button. The ROI (region of interest) box displ
7 7-50 z.one ultra and z.one ultra sp OPERATOR MANUALDuplex Mode In Duplex mode (B-Mode and PW-mode running simultaneously) an available Update opt
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-51Triplex Mode Triplex mode allows the user to visualize B-Mode, Color or Power Doppler, and PW Do
7 7-52 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 7-21. PW Autotrace Icon - Doppler Data Field3. Select the desired measurement (e.g., R
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-534. While in CW imaging, the 2D image is in Update mode. Press the Set button to update the 2D im
7 7-54 z.one ultra and z.one ultra sp OPERATOR MANUALTintsSmartCart/SmartCart sp Press Menu button. Highlight Tints. Move Trackball left/right (or
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-553D & 4D (Real-Time 3D) Imaging3D/4D imaging is available only when connected to a Multi-Tra
z.one ultra and z.one ultra sp Systemsxvi z.one ultra and z.one ultra sp OPERATOR MANUALFigure 14-30. Bent, Broken, or Missing Pins on Transducer...
7 7-56 z.one ultra and z.one ultra sp OPERATOR MANUALNOTE: See page 11-11 for more information about storing 3D Volume Data.Figure 7-23. 3D/4D Pr
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-57Figure 7-24. 3D Setup Mode Status Field5. To adjust acquisition quality (QA in the status field
7 7-58 z.one ultra and z.one ultra sp OPERATOR MANUAL9. When done, system enters 3D/4D Postprocessing mode. For OB Presets, the default 3D/4D postp
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-5914. To adjust brightness, turn B-Mode button or Brightness softkey.15. To increase/decrease the
7 7-60 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 7-28. MPR Mode - Thick SliceTomo ModeIn Tomo mode, 9 images are displayed (9-up) with
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-61Features include the following abilities:• Change line spacing between slices in mm and position
7 7-62 z.one ultra and z.one ultra sp OPERATOR MANUALRender ModeSoftkey Label: Render / MPR / Tomo Use to select the type of 3D Volume desired: Re
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-63Volume RotationSoftkey Label: Press to rotate the volume to one of 4 discrete degrees: 0 / 90
7 7-64 z.one ultra and z.one ultra sp OPERATOR MANUALElastography Elastography (or tissue elasticity imaging) is an imaging modality that detects r
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-65Figure 7-31. Elastography Default Image Display - DualTo exit Dual display and view only the El
z.one ultra and z.one ultra sp OPERATOR MANUAL 1-111Introductionz.one ultra Convertible Ultrasound SystemsThe ZONARE z.one ultra and z.one ultra sp Co
7 7-66 z.one ultra and z.one ultra sp OPERATOR MANUALRelative (Normalized) StrainAbsolute Strain
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-67Cross-Correlation AssessmentElastography User ControlsElastography user controls are accessed vi
7 7-68 z.one ultra and z.one ultra sp OPERATOR MANUAL* Indicates this control is also available in retrospective processing.Table 7-7. Elastograph
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-69CompoundingScan Engine Press Compounding softkey to toggle Compounding ON/OFF. Display Imagin
7 7-70 z.one ultra and z.one ultra sp OPERATOR MANUALFilter ✱Scan Engine Press Filter softkey to select F0, F1, or F2. Increasing Filter improves
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-71Elastography Measurements To Access Measurements for Elastography1. Freeze the desired Elastogr
7 7-72 z.one ultra and z.one ultra sp OPERATOR MANUAL To Do an Area (trace or ellipse)1. Display the Measurements menu (see “To Access Measurement
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-73Retrospective ProcessingElastography may be enabled/disabled during live, frozen, or cine imagin
7 7-74 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 7-38. Ocular Imaging Screen Contrast ImagingNOTE: For SmartCart/SmartCart sp systems
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-75NOTE: Once a Mode key is defined, it will remain so unless the system configuration is reset to
1 1-2 z.one ultra and z.one ultra sp OPERATOR MANUALIntended UseThe SmartCart/SmartCart sp Workstations are intended to be used for the following t
7 7-76 z.one ultra and z.one ultra sp OPERATOR MANUALStopwatch STOP/START Softkey Press START to start timer from 0:00:00 (hours:minutes:seconds)
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-77Control Panel ControlsDepthSmartCart/SmartCart sp Press Depth button Up to decrease depth; pres
7 7-78 z.one ultra and z.one ultra sp OPERATOR MANUALDual: OFF/ON SmartCartSmartCart spNOTE: B-Mode image displays on right and Contrast image di
7z.one ultra and z.one ultra sp OPERATOR MANUAL 7-79Maps Press Menu button. Highlight Maps with Trackball. Move Trackball Left / Right to select
7 7-80 z.one ultra and z.one ultra sp OPERATOR MANUALOptimizeSmartCart/SmartCart sp Overall / DGC Gain Press the Optimize button to automatically
z.one ultra and z.one ultra sp OPERATOR MANUAL 8-188AnnotationsThe user can enter text annotation, body pattern graphics, and arrow graphics on live i
88-2 z.one ultra and z.one ultra sp OPERATOR MANUALList Text EntriesList1, List2, and List3 menus are application-specific terms defined by the user.
8z.one ultra and z.one ultra sp OPERATOR MANUAL 8-3CheckboxesDefine Lists Click the Define Lists softkey on the Annotation Preset Configuration screen
88-4 z.one ultra and z.one ultra sp OPERATOR MANUAL To Add Item to a List1. Click on the desired List name in the List dropdown box.Figure 8-3. List
8z.one ultra and z.one ultra sp OPERATOR MANUAL 8-5 To Move Item Up or Down in List1. Highlight the Item you want to move.2. Click the Move up or dow
1z.one ultra and z.one ultra sp OPERATOR MANUAL 1-37ImagingDescribes procedures for all imaging modes.8 AnnotationsDescribes how to place annotation
88-6 z.one ultra and z.one ultra sp OPERATOR MANUAL When the cursor moves to a new position after the user enters text, the previous text string is co
8z.one ultra and z.one ultra sp OPERATOR MANUAL 8-7Overwrite Mode 1. Place the Tex t cursor between the desired characters to start Overwrite Mode.2.
88-8 z.one ultra and z.one ultra sp OPERATOR MANUALTo Enter POT from Qwerty Keyboard:1. Press the Tex t key or Spacebar on the keyboard to display the
8z.one ultra and z.one ultra sp OPERATOR MANUAL 8-93. Press the List1 key on the keyboard. The first list value displays, and the text cursor then mov
88-10 z.one ultra and z.one ultra sp OPERATOR MANUAL 3. Unfreeze the image. The System remembers the last cursor position, and then the cursor defaul
8z.one ultra and z.one ultra sp OPERATOR MANUAL 8-11Annotation Procedures - Scan Engine To Place Annotations1. From the Tools menu or the Freeze m
88-12 z.one ultra and z.one ultra sp OPERATOR MANUALNOTE: The user can enter POT values without first pressing the Tex t key. When a POT key is press
8z.one ultra and z.one ultra sp OPERATOR MANUAL 8-13Graphics Procedures - SmartCart/SmartCart spSee “Body Pattern Graphics” on page 17-9 for a complet
88-14 z.one ultra and z.one ultra sp OPERATOR MANUAL To Use ArrowsArrows can be used during live or frozen imaging. As many as 15 arrows can be dis
8z.one ultra and z.one ultra sp OPERATOR MANUAL 8-15Graphics Procedures - Scan EngineSee “Body Pattern Graphics” on page 17-9 for a complete list of
ZONARE, the ZONARE logo and z.one ultra, and Convertible Ultrasound are all trademarks of ZONARE Medical Systems Inc. All other trademarks are the pro
This page intentionally blank
8 8-16 z.one ultra and z.one ultra sp OPERATOR MANUAL1. From the Tools menu or the Freeze menu, use Menu Control to scroll to the Annotate selecti
8z.one ultra and z.one ultra sp OPERATOR MANUAL 8-17Dual Imaging Annotations - All SystemsTex t, POTS, Lists, Body Patterns, and Arrows may be enter
This page intentionally blank
z.one ultra and z.one ultra sp OPERATOR MANUAL 9-199Measurements and CalculationsNOTE: For Cardiac calculations and imaging, see Chapter 10 Echocardi
9 9-2 z.one ultra and z.one ultra sp OPERATOR MANUALTools for Measurements and CalculationsFreeze The image must be frozen before most measurements
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-3upon Freeze option is On (see Figure 9-5) and a Calculation Package is active when an un-Freeze o
9 9-4 z.one ultra and z.one ultra sp OPERATOR MANUALWhen working with Calc Packages, the data in the DDB is stored when the measurements have been
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-5Figure 9-4. Calcs Menu2. The General selection displays the General Configuration screen. Figure
9 9-6 z.one ultra and z.one ultra sp OPERATOR MANUALWorksheets and ReportsNOTE: For information on other Worksheets/Reports available on the syste
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-7 Date Data Field: The date data field accepts only a date in the format selected in system setup.
z.one ultra and z.one ultra sp OPERATOR MANUAL 2-122The Basics: SmartCart and SmartCart spz.one ultra Systems: SmartCart The z.one ultra Convertible U
9 9-8 z.one ultra and z.one ultra sp OPERATOR MANUAL2. Go to Figure 13-12, “DICOM Network Storage Destination Screen,” on page 13-10, and configure
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-9Figure 9-8. Custom Report Pages Screen3. Insert USB stick containing custom User Worksheet templ
9 9-10 z.one ultra and z.one ultra sp OPERATOR MANUAL8. Assign User Worksheet for desired Preset from Worksheet drop-down list. This list displays
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-11• Data entered in User Worksheets is saved with exam’s Calc data. Data is restored when an exam
9 9-12 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 9-10. Printer/Export Configuration Screen3. Make your selections (table below) and cli
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-13Printing OB & GYN ReportsYou can print OB & GYN Calc reports (OB & GYN trending grap
9 9-14 z.one ultra and z.one ultra sp OPERATOR MANUALExporting Report Data to Serial DeviceUsers can export Report data from OB/GYN, Abdominal, Vas
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-15 To Export Data1. Make sure the targeted Export system is connected to one of the external USB
9 9-16 z.one ultra and z.one ultra sp OPERATOR MANUALNOTE: The first item in the Measure menu is auto-selected when the user selects Measure.3. Us
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-174. Use Trackball to position caliper to desired location within image. The DDB displays Depth an
2 2-2 z.one ultra and z.one ultra sp OPERATOR MANUAL.Figure 2-1. z.one ultra Convertible Ultrasound System19” LCD DisplaySpeakersGel BottleholderC
9 9-18 z.one ultra and z.one ultra sp OPERATOR MANUALNOTE: Any gap in the outline between first and second Trace calipers will be automatically cl
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-19 To Measure DepthLive Image1. Display a live M-Mode strip image.2. Press Measure button.3. Use
9 9-20 z.one ultra and z.one ultra sp OPERATOR MANUAL6. To activate additional distance measurements (up to a maximum of four), press Enter button
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-21 To Measure Resistive Index (RI) 1. Freeze the display with a PW Doppler strip image.2. Press t
9 9-22 z.one ultra and z.one ultra sp OPERATOR MANUAL• Only a single PI measurement can be performed at any given time.NOTE: In the DDB, Max veloc
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-2311. Press the Set button to fix the PSV velocity caliper and activate the MDV velocity caliper.N
9 9-24 z.one ultra and z.one ultra sp OPERATOR MANUALNOTE: Acceleration is always calculated in the direction of positive time. The velocity calip
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-25NOTE: Pressing the Set button again toggles the active caliper selection between the first and
9 9-26 z.one ultra and z.one ultra sp OPERATOR MANUAL11. Press the Measure button to remove the A/B tools and exit Measure mode.OB Calculation Pack
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-27The following specific OB calculations are available for the Fetal Heart preset. HC MCACEREB HRB
2z.one ultra and z.one ultra sp OPERATOR MANUAL 2-3Figure 2-2. z.one ultra Ultrasound System (nonconvertible)Figure 2-3. z.one ultra sp Convertibl
9 9-28 z.one ultra and z.one ultra sp OPERATOR MANUALOB Worksheets and ReportsFigure 9-13. OB Worksheet (page 1)RV Diam [Sys/Dia] MPAIVS [Sys/Dia]
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-29Figure 9-14. OB Worksheet (page 2)Figure 9-15. OB Worksheet (page 3)
9 9-30 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 9-16. OB Worksheet (page 4)Figure 9-17. OB Worksheet (page 5)
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-31Figure 9-18. OB ReportNOTE: The actual contents and layout (including the number of pages disp
9 9-32 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 9-19. OB Worksheet - Trending PageNOTE: OB Trending requires the date for each prior
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-33OB Trending data will only be displayed on the graphs when the data are selected to be plotted f
9 9-34 z.one ultra and z.one ultra sp OPERATOR MANUAL12. To activate the first Distance caliper of the third pair, Freeze the new image and: 13. Us
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-355. Heart Rate (HR) is active by default. A green vertical bar displays on the scroll. 6. Use the
9 9-36 z.one ultra and z.one ultra sp OPERATOR MANUALAbdominal Package AortaBowelB-Mode PW DopplerAorta AP [prox/mid/dst/renl] CIA [right/left]Aort
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-37GBLiver/SpleenPancreasRenalB-Mode PW DopplerGB VolGB WallCBD [prox/mid/dst]B-Mode PW DopplerLive
2 2-4 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 2-4. z.one ultra sp Ultrasound System (nonconvertible)SmartCart/SmartCart sp FeaturesTh
9 9-38 z.one ultra and z.one ultra sp OPERATOR MANUALBladderGyn and Follicular PackageNOTE: For Follicle Diam/Vol measurements, double-clicking th
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-39Pediatric Hip Package To Acquire Pediatric Hip Angle 1. Optimize the B-Mode image.2. Freeze the
9 9-40 z.one ultra and z.one ultra sp OPERATOR MANUAL6. A second caliper tool becomes active for the next Hip Angle measurement component.Bony Roof
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-41The Worksheet To view the Worksheet: To edit the Worksheet, highlight the value to be edited, th
9 9-42 z.one ultra and z.one ultra sp OPERATOR MANUALGrafts Submenu SelectionsLower Extremity Arterial Calc Package B-Mode PW DopplerGrafts* Grafts
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-43Lower Extremity Venous Calc Package Upper Extremity Venous Calc Package 2D/B ModeNew!2D/B Mode a
9 9-44 z.one ultra and z.one ultra sp OPERATOR MANUALVenous MeasurementsFigure 9-21. Vascular Calcs Page 8* Closure (valve closure time) = reflux.
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-45Figure 9-22. Auto-Dop Trace Configuration Screen To Configure Default SettingsThe following pa
9 9-46 z.one ultra and z.one ultra sp OPERATOR MANUAL Autotrace Default • measurements - Select to automatically display Measure menu when you acti
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-47Auto ParametersThe user can override the default Auto-Dop Trace settings that were configured on
2z.one ultra and z.one ultra sp OPERATOR MANUAL 2-5Figure 2-5. SmartCart/SmartCart sp LCD DisplayFull Screen Image DisplayThe SmartCart/SmartCart s
9 9-48 z.one ultra and z.one ultra sp OPERATOR MANUALEnter, Enter/Select, or Select. Additionally, Autotrace Calc results can only be entered from
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-49Figure 9-24. Auto-Dop Trace - Automeasure (Single Cycle)Figure 9-25. Auto-Dop Trace - Automeas
9 9-50 z.one ultra and z.one ultra sp OPERATOR MANUAL4. On next menu are three selections: Vessel Diam, PW Gate Size, and Start Trace.5. Highlight
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-51Measurement AccuracyThe description of the fundamental accuracy of 2D, time, and velocity measur
9 9-52 z.one ultra and z.one ultra sp OPERATOR MANUAL The above accuracy statements are consistent with 2D measurement tests performed using a Gamm
9z.one ultra and z.one ultra sp OPERATOR MANUAL 9-53 These accuracy statements are consistent with PW elapsed time measurement tests performed usin
9 9-54 z.one ultra and z.one ultra sp OPERATOR MANUALRegistration Accuracy in CDFor CD scanning of Gammex Model 1425A Doppler Flow Phantom with a s
z.one ultra and z.one ultra sp OPERATOR MANUAL 10-11010Echocardiography OptionCardiac Calculation PackageNOTE: The Echocardiography option is availab
10 10-2 z.one ultra and z.one ultra sp OPERATOR MANUAL3. Select the options you want, then click Apply to save your changes.4. Then click the arrow
10z.one ultra and z.one ultra sp OPERATOR MANUAL 10-3The following calculations are listed in the Cardiac Report and are available for the Cardiac E
2 2-6 z.one ultra and z.one ultra sp OPERATOR MANUALVideo Level AdjustmentAll adjustments are performed from an onscreen setup menu. Available adju
10 10-4 z.one ultra and z.one ultra sp OPERATOR MANUALLV MassEpi/EndoLengthLA VolumeApical 4VolumeApical 2VolumeMitralVena ContractaMR PISA RadiusS
10z.one ultra and z.one ultra sp OPERATOR MANUAL 10-5Cardiac Imaging and Related FunctionsECG The ECG trace is available only on SmartCart systems e
10 10-6 z.one ultra and z.one ultra sp OPERATOR MANUALThe ECG trace is part of the Cardiac Exam Preset and automatically scrolls at the bottom of t
10z.one ultra and z.one ultra sp OPERATOR MANUAL 10-7Figure 10-5. ECG Softkey ControlsSweep Speed The sweep speed in B-Mode (2D) when the ECG strip
10 10-8 z.one ultra and z.one ultra sp OPERATOR MANUALM-Mode Calculations To Perform M-Mode Cardiac Calculations1. Obtain the desired M-Mode image
10z.one ultra and z.one ultra sp OPERATOR MANUAL 10-99. To edit the Worksheet, highlight the value to be edited, then press the Delete button or use
10 10-10 z.one ultra and z.one ultra sp OPERATOR MANUAL8. To view the Cardiac Worksheet, press Report.9. To edit the Worksheet, highlight the value
10z.one ultra and z.one ultra sp OPERATOR MANUAL 10-117. Repeat steps 1 - 6 for any additional measurements. 8. To view the Cardiac Worksheet, press
10 10-12 z.one ultra and z.one ultra sp OPERATOR MANUALImage Width Button (SmartCart only)The Image Width button allows you to have a narrow sect
10z.one ultra and z.one ultra sp OPERATOR MANUAL 10-13PW/CW/TDI (Pulsed Wave/Continuous Wave/Tissue Doppler Imaging) To Select Desired Doppler Mode
2z.one ultra and z.one ultra sp OPERATOR MANUAL 2-7External Video Mode The external digital video output, which is produced on the HDMI (DVI) video
10 10-14 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 10-16. PW Doppler - Doppler Data Field & LegendNOTE: For a list of the controls
10z.one ultra and z.one ultra sp OPERATOR MANUAL 10-15Figure 10-18. CW Doppler - Doppler Data Field & LegendNOTE: For a list of the controls a
This page intentionally blank
z.one ultra and z.one ultra sp OPERATOR MANUAL 11-11111Archive and Review Any screen displayed in the Image area can be saved (stored) electronically
11 11-2 z.one ultra and z.one ultra sp OPERATOR MANUALImage files cannot be retrieved from network storage for review on the System. These files ca
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-3Compression Settings The size of the stored image files is determined by the preconfigured Comp
11 11-4 z.one ultra and z.one ultra sp OPERATOR MANUAL•The JPEG Lossy method provides a significant reduction in file size (more compression than R
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-5 To Store to FTPYou can store exam data/images directly to an FTP site. See “FTP Storage” on p
11 11-6 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 11-4. Image Store/Print Buttons Screen* To print (by pressing Print or Store button)
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-7Live Capture Options: Clip: If ultrasound image is live (Freeze button not depressed), pressin
2 2-8 z.one ultra and z.one ultra sp OPERATOR MANUALIMPORTANT: Before deleting any Exam data from the Scan Engine or Cart hard drive, always verify
11 11-8 z.one ultra and z.one ultra sp OPERATOR MANUAL To Configure Store/Print ButtonNOTE: See also “Setting Up or Adding a Local (USB) Printer”
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-9Still & Clip Image StoreThe Store and/or Print buttons can be configured to send the curren
11 11-10 z.one ultra and z.one ultra sp OPERATOR MANUALIQ Scan DataThe IQ Scan feature allows you to store raw source (prescan converted/processed)
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-116. While IQ Data is playing back, you can store a clip with the new parameter settings.NOTE:
11 11-12 z.one ultra and z.one ultra sp OPERATOR MANUALDICOM Transfer of Clip Image StoresStore and Print buttons may be configured to send clip im
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-13• Press Back key twice. The first press displays Current Exam review menu; the second press re
11 11-14 z.one ultra and z.one ultra sp OPERATOR MANUAL To Play a Clip Image From Review1. Use Pg Fwd/Back or Trackball to find the desired Clip (
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-15c. Place the pointer to the left of the desired parameter(s) on the pull-down menu, and press
11 11-16 z.one ultra and z.one ultra sp OPERATOR MANUALa. With the wrong exam still highlighted, move pointer over desired exam and press Set butto
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-173. If the wrong exam is selected:a. With the wrong exam still highlighted, move the pointer ov
2z.one ultra and z.one ultra sp OPERATOR MANUAL 2-9Docking/Undocking Scan Engine/Scan ModuleThe Scan Engine/Scan Module must be docked on the SmartC
1111-18 z.one ultra and z.one ultra sp OPERATOR MANUAL 11. The images are deleted and the System returns to live imaging. If the user opens a current
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-197. Multiple full-size images can be tagged for deletion by using the Pg Back and Pg Fwd keys t
11 11-20 z.one ultra and z.one ultra sp OPERATOR MANUALStoring Archived Exams to DICOM DevicesSmartCart/SmartCart sp Systems can store archived exa
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-216. The Patient Information screen displays, populated with patient demographic data from resta
11 11-22 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 11-9. Archive Exam Export Screen3. In Destination field, select Cart HD and press Se
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-231. Insert a blank disc into drive. Disc icon will appear in upper left corner of imaging scre
11 11-24 z.one ultra and z.one ultra sp OPERATOR MANUAL To Customize Archive Exam Export Options1. On Archive Exam Export screen (Figure 11-9), us
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-25Remove identity of patient(s) To remove the identity of patient(s), check the box next to Re
11 11-26 z.one ultra and z.one ultra sp OPERATOR MANUAL4. Click Apply to save settings.NOTE: When the patient identity is removed from exported ex
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-277. Select Yes if you want only keyed images to export.8. Use Export feature as described above
z.one ultra and z.one ultra sp Systemsz.one ultra and z.one ultra sp OPERATOR MANUAL iiiContentsContents...
2 2-10 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 2-9. SmartCart/SmartCart sp Keylock LocationPowering ON/OFFPlease read and understand
11 11-28 z.one ultra and z.one ultra sp OPERATOR MANUAL To View Studies on Exported CD/DVD1. Insert the disc into the appropriate drive on your PC
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-29 To view an image or clip in larger format, double-click it. An enlarged image appears that ca
11 11-30 z.one ultra and z.one ultra sp OPERATOR MANUAL Click to display the Patient/Study Information screen:Clip Slider When a clip is stopped, y
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-31Enlarged Image/Clip ScreenImages/clips can be adjusted (contrast/brightness), annotated, and e
11 11-32 z.one ultra and z.one ultra sp OPERATOR MANUALAVI CompressionAVI files are extremely large when uncompressed, so you want to compress clip
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-33 To Export Clips and Set AVI Compression1. View the study in the Series window. 2. Highlight
11 11-34 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 11-17. Compressor Options Drop-Down Box5. To set Compression Quality (Figure 11-16),
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-35Figure 11-18. Archive Exam Import Screen4. Use Trackball and Set button to highlight desired
11 11-36 z.one ultra and z.one ultra sp OPERATOR MANUALPatient Exam Management ScreenOn the Patient Exam Management screen, users can perform tasks
11z.one ultra and z.one ultra sp OPERATOR MANUAL 11-37Patient exam database rebuild helper files Only applies to patient exams on Scanner SSD or C
2z.one ultra and z.one ultra sp OPERATOR MANUAL 2-112. The System will take approximately 30 seconds to complete the normal power on sequence.3. Che
This page intentionally blankEmergency System Backup To export all internally stored patient exam archive to a USB Memory Stick because the System is
z.one ultra and z.one ultra sp OPERATOR MANUAL 12-11212Printers and PeripheralsNOTE: For information and pricing on ZONARE system upgrades & feat
12 12-2 z.one ultra and z.one ultra sp OPERATOR MANUALFollow guidelines provided by IEC 60601 when connecting peripherals (e.g., VCR).Figure 12-1.
12z.one ultra and z.one ultra sp OPERATOR MANUAL 12-3Status LED The status LED (Figure 12-2) displays the current status of I/O activities as follow
12 12-4 z.one ultra and z.one ultra sp OPERATOR MANUALThe following section assumes that the Sony UP-D897 B&W USB Printer has already been moun
12z.one ultra and z.one ultra sp OPERATOR MANUAL 12-5NOTE: Due to its physical size, the Sony UP-D25MD printer is not mountable within the chassis
12 12-6 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 12-5. Printer Screen3. Using the pull-down menu under the Type selection, select Loca
12z.one ultra and z.one ultra sp OPERATOR MANUAL 12-7Figure 12-7. Image Store/Print Buttons Screen: Print Section2. In the Print section of the scr
This page intentionally blank
z.one ultra and z.one ultra sp OPERATOR MANUAL 13-11313DICOM and FTP ConnectivityDICOM connectivity is an optional software feature on SmartCart/Smart
2 2-12 z.one ultra and z.one ultra sp OPERATOR MANUALHeight Adjustment To Adjust Height of User Interface Panel1. Squeeze the release lever locate
13 13-2 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 13-1. Network Configuration Screen3. Fill in the data fields (see Table 13-1). Use th
13z.one ultra and z.one ultra sp OPERATOR MANUAL 13-3Wireless Network ConnectivityZONARE supports wireless connectivity using the Quatech AirborneDi
13 13-4 z.one ultra and z.one ultra sp OPERATOR MANUALLocationThe Location feature allows users to quickly activate network/DICOM functionality whe
13z.one ultra and z.one ultra sp OPERATOR MANUAL 13-54. Click Apply to save the new Location Profile. 5. The Location name now appears in the Locati
13 13-6 z.one ultra and z.one ultra sp OPERATOR MANUALGeneral To Configure General DICOM Parameters1. Make sure the System has a live Ethernet con
13z.one ultra and z.one ultra sp OPERATOR MANUAL 13-7PrintersOnly DICOM-compliant printers can be configured using this procedure. Consult a network
13 13-8 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 13-9. Printer Administration Screen4. Select New or Edit. The Printer screen displays
13z.one ultra and z.one ultra sp OPERATOR MANUAL 13-9 If a low-level network connection has been made, the following message appears:ICMP Echo: xx.
13 13-10 z.one ultra and z.one ultra sp OPERATOR MANUALNetwork Storage To Configure a DICOM Network Storage Destination1. Make sure the System has
13z.one ultra and z.one ultra sp OPERATOR MANUAL 13-115. Fill in the data fields (see Table 13-4). Use the QWERTY Tab key (or Trackball and Set butt
2z.one ultra and z.one ultra sp OPERATOR MANUAL 2-13Figure 2-14. Wheel Locking/Brake Pedal (SmartCart/SmartCart sp)Moving the SmartCart/SmartCart s
13 13-12 z.one ultra and z.one ultra sp OPERATOR MANUALColor Model(Photometric Interpretation)Method to be used by the DICOM receiving device for i
13z.one ultra and z.one ultra sp OPERATOR MANUAL 13-13Reconciliation For exams not yet sent to PACS. Makes the patient/study data in all images in a
13 13-14 z.one ultra and z.one ultra sp OPERATOR MANUALDICOM Structured ReportsIMPORTANT: The configuration required for storing DICOM Structured
13z.one ultra and z.one ultra sp OPERATOR MANUAL 13-15Figure 13-13. DICOM MPPS Administration Screen4. Select New or Edit. The DICOM MPPS Destinati
13 13-16 z.one ultra and z.one ultra sp OPERATOR MANUALICMP Echo: xx.xx.xx.xx timed out [xx.xx.xx.xx is the IP address of the configured device].b.
13z.one ultra and z.one ultra sp OPERATOR MANUAL 13-17Go to: Setup button | System Setup | Archive | DICOM | Worklist2. The DICOM Modality Worklist
13 13-18 z.one ultra and z.one ultra sp OPERATOR MANUAL If a low-level network connection has been made, the following message appears:ICMP Echo:
13z.one ultra and z.one ultra sp OPERATOR MANUAL 13-19Store/Print ButtonsThe Store and Print buttons on the SmartCart/SmartCart sp and Scan Engine m
13 13-20 z.one ultra and z.one ultra sp OPERATOR MANUAL To Configure Store/Print Buttons for DICOM Printing1. Locate the appropriate button config
13z.one ultra and z.one ultra sp OPERATOR MANUAL 13-21FTP Storage Setup1. Go to System Setup | Archive | FTP | FTP Store. The FTP Storage Administr
2 2-14 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 2-16. Cable HooksCAUTION: To avoid possible damage to transducer cables during moveme
13 13-22 z.one ultra and z.one ultra sp OPERATOR MANUAL4. Fill in the data fields (Table 13-7) and click Apply. The system returns to the FTP Stora
13z.one ultra and z.one ultra sp OPERATOR MANUAL 13-23Quality Move slider to select compression quality (see “Compression Settings” on page 11-3).
This page intentionally blank
z.one ultra and z.one ultra sp OPERATOR MANUAL 14-11414User MaintenanceThis chapter describes the maintenance, care, and service that should be perfor
14 14-2 z.one ultra and z.one ultra sp OPERATOR MANUALDo not use any transducer that has been immersed beyond the specified maximum immersion depth
14z.one ultra and z.one ultra sp OPERATOR MANUAL 14-3Touchscreen Calibration Procedureu To Calibrate the Touchscreen1. Undock the Scan Engine from t
14 14-4 z.one ultra and z.one ultra sp OPERATOR MANUAL• Capturing of system status to LOG files• Capturing current image screen and storing as a *.
14z.one ultra and z.one ultra sp OPERATOR MANUAL 14-5Settings To Access the Settings Screen1. Click Settings on the User Diagnostics Panel (Figur
14 14-6 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 14-5. ZPak Battery Status Message; Click OK to closePreset Mgmt/System Setup: Backup/
14z.one ultra and z.one ultra sp OPERATOR MANUAL 14-74. Press Setup button. The Setup menu displays:Figure 14-6. Setup Button Menu5. Highlight and
2z.one ultra and z.one ultra sp OPERATOR MANUAL 2-15USB Memory SticksThe SmartCart/SmartCart sp uses removable media (e.g., USB Memory Sticks) for t
14 14-8 z.one ultra and z.one ultra sp OPERATOR MANUAL To Restore Presets1. Highlight and select Restore Presets (Figure 14-7). The following menu
14z.one ultra and z.one ultra sp OPERATOR MANUAL 14-9c. System displays as Restore proceeds.3. Select Exam Type (Figure 14-10) to individually sel
14 14-10 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 14-16. Restore Presets | Single Preset Screena. Select Transducer, exam type (Catego
14z.one ultra and z.one ultra sp OPERATOR MANUAL 14-11Figure 14-18. Backup Configuration Screen7. All options are checked (default). Click Backup t
14 14-12 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 14-20. System Restore Configuration Screen7. All options are checked (default). Clic
14z.one ultra and z.one ultra sp OPERATOR MANUAL 14-13Figure 14-22. Restore Factory Menu5. Select All to restore all factory Presets. The following
14 14-14 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 14-26. Preset Settings On/Off Screenc. Click Ye s to keep your presets visible in s
14z.one ultra and z.one ultra sp OPERATOR MANUAL 14-15• 1-650-316-3199Verifying Current System Software Revision To Verify Current Software Revisio
14 14-16 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 14-29. Insert Removable Media6. A software installation dialog box will display on t
14z.one ultra and z.one ultra sp OPERATOR MANUAL 14-17Install media detected. Do you wish to install new software: Yes/No5. To begin the software in
This page intentionally blank
14 14-18 z.one ultra and z.one ultra sp OPERATOR MANUAL3. Press the Power button on the Scan Engine, and then immediately make note of the battery
14z.one ultra and z.one ultra sp OPERATOR MANUAL 14-19A Low Storage warning will appear when “free” storage falls below a minimum threshold. When yo
14 14-20 z.one ultra and z.one ultra sp OPERATOR MANUALWARNING: Do not expose the external case of the LCD Display to any of the following agents:
14z.one ultra and z.one ultra sp OPERATOR MANUAL 14-21WARNING: Bent, broken, or missing pins on the transducer connector may cause poor image quali
14 14-22 z.one ultra and z.one ultra sp OPERATOR MANUALSystem Mechanical & Visual InspectionVerifying System Display1. Verify the Display is se
14z.one ultra and z.one ultra sp OPERATOR MANUAL 14-23Verifying PeripheralsB/W and Color Printer1. Acquire the desired ultrasound image on the Smart
14 14-24 z.one ultra and z.one ultra sp OPERATOR MANUALWhen the SmartCart/SmartCart sp is not in use, be certain to leave the power cord connected
z.one ultra and z.one ultra sp OPERATOR MANUAL 15-11515TroubleshootingZONARE Contact InformationZONARE Medical Systems Inc. 420 N. Bernardo Avenue Mou
15 15-2 z.one ultra and z.one ultra sp OPERATOR MANUALSubsidiariesUser Troubleshooting ProceduresBefore calling ZONARE Technical Support, please pe
15z.one ultra and z.one ultra sp OPERATOR MANUAL 15-3 Image too soft or hazy • Increase Edge setting• Decrease Persistence setting Image too much
z.one ultra and z.one ultra sp OPERATOR MANUAL 3-133Imaging Controls: SmartCart and SmartCart spControl PanelsFigure 3-1. SmartCart Control Panel 123
15 15-4 z.one ultra and z.one ultra sp OPERATOR MANUAL• In the onscreen menu under the Preset tab, select one of the factory default settings (Gene
15z.one ultra and z.one ultra sp OPERATOR MANUAL 15-5• Check the placement of the ECG pads on the patient and ensure connectivity of all leads.• Ins
15 15-6 z.one ultra and z.one ultra sp OPERATOR MANUALEnhancement is an increased relative amplitude of echoes caused by an intervening structure o
15z.one ultra and z.one ultra sp OPERATOR MANUAL 15-7Side lobes (from single-element transducers) and grating lobes (from array transducers) cause o
15 15-8 z.one ultra and z.one ultra sp OPERATOR MANUAL• If either component fails to start-up after these steps, contact ZONARE Technical Support.
15z.one ultra and z.one ultra sp OPERATOR MANUAL 15-9Printer Error MessagesTransducer Problems Transducer Not Recognized By System (No B-Mode Imagi
15 15-10 z.one ultra and z.one ultra sp OPERATOR MANUAL• If problem continues, call ZONARE Technical Support to verify DICOM option has been purcha
z.one ultra and z.one ultra sp OPERATOR MANUAL 16-11616SpecificationsThis chapter provides electrical, connection, physical, environmental, and author
16 16-2 z.one ultra and z.one ultra sp OPERATOR MANUALPhysical (weight)SmartCart 65.3 kg (144 lb)Scan Engine – (w/battery) 2.5 kg (5.6 lb)SmartCar
16z.one ultra and z.one ultra sp OPERATOR MANUAL 16-3Power consumption (w/peripherals) (max.)442W (1512 BTU/Hr)Environmental (operating)Cooling requ
3 3-2 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 3-2. SmartCart sp Control PanelTable 3-1. SmartCart/SmartCart sp Control Panels: Syste
16 16-4 z.one ultra and z.one ultra sp OPERATOR MANUALI/O connectors Ethernet (1 port) RJ-45 – 10/100BaseTUSB 2.0 (4 ports) USB-Type A (VBUS 5.0V,
16z.one ultra and z.one ultra sp OPERATOR MANUAL 16-5Width 58 mm (2.3 in.)Depth 132 mm (5.2 in.)Table 16-4. Specifications - AC Power Adapter Type
16 16-6 z.one ultra and z.one ultra sp OPERATOR MANUALWidth 260 mm (10 in)Depth 355 mm (14 in)Table 16-5. Specifications - SmartCart Battery Pack
z.one ultra and z.one ultra sp OPERATOR MANUAL 17-11717Glossary, Abbreviations, and Body Pattern GraphicsThis chapter contains a glossary, a list of a
17 17-2 z.one ultra and z.one ultra sp OPERATOR MANUALCurved Array Transducer Identified by the letter C (curved) and a pair of numbers (6-2) = e.g
17z.one ultra and z.one ultra sp OPERATOR MANUAL 17-3PAL Phase Alternating Line: A video standard format setting used in a number of different count
17 17-4 z.one ultra and z.one ultra sp OPERATOR MANUALwhen excited electrically, emit acoustic energy. When the acoustic energy is transmitted into
17z.one ultra and z.one ultra sp OPERATOR MANUAL 17-5BOD biocular distance (diameter); aka IOD (interocular distance - distance between the eyes)BX
17 17-6 z.one ultra and z.one ultra sp OPERATOR MANUALEV endovaginalFA femoral arteryFV femoral veinFc center frequencyFIB fibulaFL femur lengthGS
17z.one ultra and z.one ultra sp OPERATOR MANUAL 17-7MPA main pulmonary arteryMPPS modality performed procedure stepMR mitral regurgitationMR Pisa m
3z.one ultra and z.one ultra sp OPERATOR MANUAL 3-38Exam Type ButtonPress to display the Exam Type/Presets menu and 6 Preset softkeys configured for
17 17-8 z.one ultra and z.one ultra sp OPERATOR MANUALRad V radial veinRI resistance indexROI region of interestRV right ventricleRVIDd right ventr
17z.one ultra and z.one ultra sp OPERATOR MANUAL 17-9Body Pattern GraphicsBody Pattern Graphic Exam Type(s) Body Pattern GraphicExam Type(s)abd_gbAb
17 17-10 z.one ultra and z.one ultra sp OPERATOR MANUALbody_arm_palmVascularob_breechOBbody_gynOBGynecologyob_breech_breechOBbody_head_postPediatri
17z.one ultra and z.one ultra sp OPERATOR MANUAL 17-11body_lldAbdomenPediatricsp_testiclesSmall Parts body_lpo & _rpoAbdomenPediatricob_LT_RTOB
17 17-12 z.one ultra and z.one ultra sp OPERATOR MANUALbody_supineAbdomenOBGynecologySmall PartsPediatric vasc_carotid LT/RTVascularBody Pattern Gr
z.one ultra and z.one ultra sp OPERATOR MANUAL Index-1z.one ultra and z.one ultra sp SystemsIndexNumerics3D imaging controls 7-613D MPR mode 7-593D po
z.one ultra and z.one ultra sp SystemsIndex-2 z.one ultra and z.one ultra sp OPERATOR MANUALbarcode reader 2-8, 3-22baseline 7-39, 7-45, 7-53batterysc
In-z.one ultra and z.one ultra sp OPERATOR MANUAL Index-3z.one ultra and z.one ultra sp Systemsdual mode 7-36edge 7-41filter 7-39flash cancel 7-40freq
z.one ultra and z.one ultra sp SystemsIndex-4 z.one ultra and z.one ultra sp OPERATOR MANUALupdate option 7-50dynamic range 7-24, 7-33, 7-47, 7-54EE/A
In-z.one ultra and z.one ultra sp OPERATOR MANUAL Index-5z.one ultra and z.one ultra sp Systemsimage screenSmartCart/SmartCart sp 3-7image store/print
z.one ultra and z.one ultra sp Systemsiv z.one ultra and z.one ultra sp OPERATOR MANUALImaging Screen...
3 3-4 z.one ultra and z.one ultra sp OPERATOR MANUALSmartCart/SmartCart sp KeyboardFigure 3-3. SmartCart/SmartCart sp Keyboard29Annotation ButtonP
z.one ultra and z.one ultra sp SystemsIndex-6 z.one ultra and z.one ultra sp OPERATOR MANUALautomeasure 9-47pulsatility index 9-21resistivity index 9-
In-z.one ultra and z.one ultra sp OPERATOR MANUAL Index-7z.one ultra and z.one ultra sp Systemsprinter screen 12-6, 13-8printer/export configuration s
z.one ultra and z.one ultra sp SystemsIndex-8 z.one ultra and z.one ultra sp OPERATOR MANUALupgrade 14-14verify current revision 14-15Sony UP-D23MD Di
In-z.one ultra and z.one ultra sp OPERATOR MANUAL Index-9z.one ultra and z.one ultra sp SystemsP4-1 6-24P4-1c 6-24P8-3TEE 6-24P9-3ic 6-24protective sh
This page intentionally blank
3z.one ultra and z.one ultra sp OPERATOR MANUAL 3-54 PG FWD Same as PG BK, except the image advances in direction.5 FIRST LAST This key only works w
3 3-6 z.one ultra and z.one ultra sp OPERATOR MANUAL14 BODY PATTERN This is a toggle key. Pressing the Body Pattern key displays a Body Pattern for
3z.one ultra and z.one ultra sp OPERATOR MANUAL 3-7Imaging ScreenFigure 3-4. SmartCart/SmartCart sp Image ScreenNew! ZPak Battery Charging Status
3 3-8 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 3-5. Image Screen - Description of LayoutStatus Icons Table 3-3 shows all status icons
3z.one ultra and z.one ultra sp OPERATOR MANUAL 3-9Scan Engine/Module Battery Battery status is unknownSmartCart/SmartCart sp BatteryBattery is full
3 3-10 z.one ultra and z.one ultra sp OPERATOR MANUALScan Engine/Scan Module Internal MediaImage storage is occurring to the internal mediaScan Eng
3z.one ultra and z.one ultra sp OPERATOR MANUAL 3-11SmartCart/SmartCart sp CD/DVD Disc (option)CD/DVD disc is inserted and ZONARE system is writing
3 3-12 z.one ultra and z.one ultra sp OPERATOR MANUALNon-Imaging StatusExam ClockThe exam clock shows:• Current time of day (analog clock)• Duratio
3z.one ultra and z.one ultra sp OPERATOR MANUAL 3-13• Default image/clip storage space insufficient for transferRecent MeasurementsRecent Measuremen
1z.one ultra and z.one ultra sp Systemsz.one ultra and z.one ultra sp OPERATOR MANUAL vExamination ...
3 3-14 z.one ultra and z.one ultra sp OPERATOR MANUALDepartment ScheduleDepartment Schedule provides quick reference to the ultrasound examinations
3z.one ultra and z.one ultra sp OPERATOR MANUAL 3-15 Table 3-4. Frequently Used Menus Button Button Name MenuSetup or Set-upMenu that appears when
3 3-16 z.one ultra and z.one ultra sp OPERATOR MANUAL• After you push one of the buttons listed in Table 3-4, its menu appears on the LCD Display.
3z.one ultra and z.one ultra sp OPERATOR MANUAL 3-17 To Navigate System MenusNOTE: All users must be familiar with the following procedure. The Tr
3 3-18 z.one ultra and z.one ultra sp OPERATOR MANUALTrackball SensitivitySmartCart/SmartCart sp users can adjust the sensitivity of the Trackball
3z.one ultra and z.one ultra sp OPERATOR MANUAL 3-19NOTE: You may have to adjust Trackball sensitivity a few times before you find the settings you
3 3-20 z.one ultra and z.one ultra sp OPERATOR MANUALSet Time, Date, and Language To Set Time, Date, and Language1. Go to: 2. The Date & Time
3z.one ultra and z.one ultra sp OPERATOR MANUAL 3-21Sleep and Shutdown To Set Sleep and Shutdown Parameters1. Go to: 2. Select the parameters you w
This page intentionally blankCAUTION: If password protection (Access Control) is used, make sure user-assigned passwords are retained in a safe place
3z.one ultra and z.one ultra sp OPERATOR MANUAL 3-233. Press New Patient key on QWERTY keyboard to display Patient Information form (Figure 7-1 on p
z.one ultra and z.one ultra sp Systemsvi z.one ultra and z.one ultra sp OPERATOR MANUALPostprocessing Modes...
This page intentionally blank
z.one ultra and z.one ultra sp OPERATOR MANUAL 4-144The Basics: Scan Engine and Scan Module Scan Engine Basics...
4 4-2 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 4-1. Scan Engine with Transducer (Front View)Figure 4-2. Scan Engine (Back View)The Sc
4z.one ultra and z.one ultra sp OPERATOR MANUAL 4-3•The Scan Engine's internal battery is automatically recharged while the Scan Engine is dock
4 4-4 z.one ultra and z.one ultra sp OPERATOR MANUALUSB Memory SticksThe Scan Engine/Scan Module uses removable media (e.g., USB Memory sticks) for
4z.one ultra and z.one ultra sp OPERATOR MANUAL 4-5shows whether the battery is charging and how much charge is left. See Table 4-3 on page 4-9. If
4 4-6 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 4-5. Scan Engine Onscreen Battery Charge Icon2. Four battery charge LEDs on the front p
4z.one ultra and z.one ultra sp OPERATOR MANUAL 4-7Scan Engine Battery Level MessagesWARNING: Battery Thermal Protection To protect the Scan Engine
4 4-8 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 4-6. Scan Engine Battery Charger To Charge a BatteryRefer to Figure 4-6 as you read th
4z.one ultra and z.one ultra sp OPERATOR MANUAL 4-9NOTE: Recalibration may take 14 hours. Do not remove the battery before recalibration is complet
1z.one ultra and z.one ultra sp Systemsz.one ultra and z.one ultra sp OPERATOR MANUAL viiEnding & Restoring Exams...
4 4-10 z.one ultra and z.one ultra sp OPERATOR MANUAL To Connect the AC Power AdapterMake sure the Scan Engine is undocked from the SmartCart/Smar
z.one ultra and z.one ultra sp OPERATOR MANUAL 5-155Imaging Controls: Scan EngineScan Engine LCD DisplayFigure 5-1. Scan Engine LCD DisplayThe Scan E
5 5-2 z.one ultra and z.one ultra sp OPERATOR MANUALVirtual Keyboard (VKB)A touchscreen keyboard, called the virtual keyboard (VKB), can be display
5z.one ultra and z.one ultra sp OPERATOR MANUAL 5-3User ControlsScan Engine Control PanelFigure 5-3. Scan Engine Control PanelTable 5-1. Scan Engi
5 5-4 z.one ultra and z.one ultra sp OPERATOR MANUAL5, 16STORE Stores current still image, or clip store sequence, to device(s) previously specifie
5z.one ultra and z.one ultra sp OPERATOR MANUAL 5-5System Navigation To View Main System Menu NOTE: All users must be familiar with the following
5 5-6 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 5-5. Scan Engine Tools MenuNOTE: In some cases, a submenu will appear. System Setup, f
5z.one ultra and z.one ultra sp OPERATOR MANUAL 5-76. For ON/OFF selections (Figure 5-7), highlight the item using the Menu Control up/down arrow ke
5 5-8 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 5-8. Presets Tab MenuPatient Tab The Patient tab allows access to patient demographic i
z.one ultra and z.one ultra sp OPERATOR MANUAL 6-166Transducers and MTP OptionZONARE Transducers...
z.one ultra and z.one ultra sp Systemsviii z.one ultra and z.one ultra sp OPERATOR MANUALDoppler Calculations ...
6 6-2 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 6-1. TransducerConnecting/DisconnectingAll ZONARE transducers connect directly to the S
6z.one ultra and z.one ultra sp OPERATOR MANUAL 6-3WARNING: Bent, broken, or missing pins on the transducer connector may cause poor image quality,
6 6-4 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 6-4. Multi-Transducer Port (attached to Scan Engine) To Connect MTP to Scan Engine/Sca
6z.one ultra and z.one ultra sp OPERATOR MANUAL 6-5Transducer Protective Sheath To Apply a Transducer Protective SheathIMPORTANT: For the protectio
6 6-6 z.one ultra and z.one ultra sp OPERATOR MANUAL3. Attach the sterile needle guide over the transducer cover.4. To activate the onscreen needle
6z.one ultra and z.one ultra sp OPERATOR MANUAL 6-7Figure 6-6. P8-3TEE TransducerIntended Use The physician conducting the examination must exercis
6 6-8 z.one ultra and z.one ultra sp OPERATOR MANUALWARNING: Any visible evidence of damage or improper mechanical operation indicates the transdu
6z.one ultra and z.one ultra sp OPERATOR MANUAL 6-9CAUTION: Devices containing latex may cause severe allergic reaction in latex sensitive individu
6 6-10 z.one ultra and z.one ultra sp OPERATOR MANUALTip Deflection Control The deflection control wheels found on the handle of the transducer con
6z.one ultra and z.one ultra sp OPERATOR MANUAL 6-11Figure 6-9. Friction Lock Operation - Anterior/PosteriorTransducer Scan Plane Rotation ControlT
1z.one ultra and z.one ultra sp Systemsz.one ultra and z.one ultra sp OPERATOR MANUAL ixDICOM Structured Reports ...
6 6-12 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 6-11. Scan Plane IndicatorTemperature IndicatorThe transducer has a temperature indica
6z.one ultra and z.one ultra sp OPERATOR MANUAL 6-13• To lower temperature, freeze image for a short time or lower Acoustic Output power3. MAXIMUM T
6 6-14 z.one ultra and z.one ultra sp OPERATOR MANUAL Prolonged pressure on the esophagus by the tip of the transducer may lead to a pressure necro
6z.one ultra and z.one ultra sp OPERATOR MANUAL 6-15Figure 6-17. TEE Transducer - Connector and PinsWARNING: Immediately replace a transducer that
6 6-16 z.one ultra and z.one ultra sp OPERATOR MANUALCleaning Wipe Method1. Disconnect the transducer from the Cart. If present, remove the trans
6z.one ultra and z.one ultra sp OPERATOR MANUAL 6-17NOTE: Consult the manufacturer’s Instructions For Use before using a TEE cleaner or disinfectan
6 6-18 z.one ultra and z.one ultra sp OPERATOR MANUALFigure 6-18. Typical Leakage Tester Setup Leakage Test Setup1. Fill soak tray or container t
6z.one ultra and z.one ultra sp OPERATOR MANUAL 6-19Figure 6-19. Leakage TesterNOTE: The leakage tester comes already programmed from ZONARE. Plea
6 6-20 z.one ultra and z.one ultra sp OPERATOR MANUALStoring the TEE Transducer Use great care when storing the TEE transducer. Make sure the tran
6z.one ultra and z.one ultra sp OPERATOR MANUAL 6-21Transducer ApplicationsNOTE: ZONARE transducers have a specific range of acceptable application