7
z.one ultra and z.one ultra sp OPERATOR MANUAL 7-61
Features include the following abilities:
• Change line spacing between slices in mm and position of the set of lines
using the Trackball
• Select reference image (one only) using softkey
• Toggle between One-up and 9-up
• Perform measurements on One-up image (Figure 7-30)
• Toggle slice lines' orientation from vertical to horizontal via softkey
Figure 7-30. Tomo Mode - One-Up Display with Measurements
3D/4D Imaging
Controls
Applicable only when connected to a Multi-Transducer Port (MTP) on SmartCart/
SmartCart sp
systems.
Table 7-6. 3D/4D Controls - Volume Acquisition Processing
Control SmartCart/SmartCart sp
Preprocessing
3D Mode
User-assigned Mode Key or Function
key
■ Press to place the 3D ROI on the image
■ Press Set to toggle between Position and Size
Preprocessing - Softkeys (see Figure 7-23 on page 7-56)
Start / Stop
Softkey Label: Start
■ Press to invoke the start of a 3D Volume
■ Pressing again during acquisition will stop it
■ Press the Freeze button to do the same thing
Commentaires sur ces manuels