7-26 z.one ultra and z.one ultra sp OPERATOR MANUAL
Compounding✱
SmartCart/
SmartCart sp
SmartCart
NOTE: All ZONARE ultrasound factory Presets have Compounding as the default. To adjust
B-Mode frequencies, first exit Compounding as follows:
■ Highlight Compounding in the Imaging menu. Press Select to turn OFF the Compound
Harmonic imaging function. Repeat this process to turn ON Compounding.
■ There may be more than one Compounding Image selection depending on the
transducer. To select another Compounding setting, press the left/right arrows on the
Menu Control.
NOTE: On the SmartCart, press the Frequency button up or down when doing Compound
imaging to cycle through Frequency and Compounding choices.
B Steer
NOTE: For linear array transducers only
■ Turn B Steer softkey to select: B Steer (0°); B Steer 10 (10° right); or B Steer –10 (10°
left). Note: L8-3 transducer B Steer options: B Steer (0°), B Steer 15 (15° right), B Steer –
15 (15° left). See also
“Needle Viz Preset (Linear array transducers)” on page 7-20.
A Output SmartCart/SmartCart sp
■ Press the Menu button. Highlight A Output with the Trackball. Move the Trackball left/
right (or press left/right Set buttons) to decrease or increase the desired transmit power
output level.
■ The current transmit power level is displayed (as a percentage; 25%, 50%, etc.) in the
Image Information area. The default value is 100%.
Biopsy Guide
SmartCart/
SmartCart sp
Scan Engine
■ From the QWERTY keyboard, press Bx
Guide key to toggle through the available
guide lines, and biopsy guide on/off, in live
or freeze modes.
■ Highlight Bx Guide. Press Select to toggle
biopsy guide on/off.
Table 7-1. B-Mode Controls (Continued)
B Control
SmartCart/SmartCart sp Scan Engine
✱ Indicates this control is also available in retrospective processing.