z.one ultra and z.one ultra sp OPERATOR MANUAL 6-3
WARNING: Bent, broken, or missing pins on the transducer connector may cause
poor image quality, including possible mirror image artifact. Be sure to check pins
before connecting transducer to the
ZONARE ultrasound system. If pins are bent,
broken, or missing, do not use the transducer and call
ZONARE Technical
Support.
Figure 6-3. Bent, Broken, or Missing Pins on Transducer
2. Insert the Transducer Connector into the Scan Engine/Scan Module’s
Transducer Connector Port. Ensure it is fully seated.
3. Turn the Locking Knob on the Transducer Connector one-quarter turn
clockwise to lock the transducer in place.
4. If properly connected to the System, the LCD monitor will display the
message Initializing transducer.
5. After ~10 seconds, a basic 2D (B-mode) image will display on the screen. The
transducer is ready for scanning.
To Disconnect a ZONARE Transducer
1. Turn the Locking Knob on the Transducer Connector one-quarter turn
counterclockwise.
2. Remove the Transducer Connector from the Transducer Connector Port.
3. The message No transducer connected will display on the screen if user is
in live imaging mode.
NOTE: To store the transducer, place it in the transducer connection holders on the
SmartCart/SmartCart sp.
Multi-Transducer Port Option
SmartCart/SmartCart sp systems with the optional Multi-Transducer Port (MTP)
are able to simultaneously connect up to 3 separate transducers at one time.
Bent pins
Missing/
damaged
pin