6-14 z.one ultra and z.one ultra sp OPERATOR MANUAL
■
Prolonged pressure on the esophagus by the tip of the transducer may lead to
a pressure necrosis phenomenon. Thus, in monitoring applications, the tip
should be removed from the esophagus wall when you are not scanning by
releasing it in the neutral position. If continuous monitoring is required, the
transducer tip should often be repositioned.
■
Whenever active scanning is not desired 'freeze' the image and release the
deflection control.
■
Position the array so that good acoustical contact is made. Before rotating the
array, actuate the friction ('brake') on the up/down movement. If this is not
done, rotating the array can mean loss of acoustical contact.
TEE – Care and Maintenance
NOTE: The following sections are for TEE transducer maintenance only.
See “Transducer Maintenance” on page 14-20 for information on maintaining,
cleaning, and disinfecting all other ZONARE transducers.
Inspecting
Transducer
WARNING: Regularly check transducer for signs of wear or damage. Inspect
transducers at least weekly for the following problems:
■
Cracks in case or transducer face.
■
Cuts or gouges on any part of the transducer, case, cable, and connector.
■
Cracks in the lens material on the transducer face.
■
Any cracks or damage to the transducer connector.
■
Proper mechanical operation of the transducer. Rotate the scan plane with the
push-buttons on the handle. Make sure that the rotation is smooth for all scan
planes. Deflect the tip in both anterior/posterior and left/right direction using
the deflection control wheel on the handle. Make sure the deflection operates
smoothly.
■
Anterior/posterior and/or left/right deflection, which may, after prolonged use,
develop an unwanted amount of free play. In that case, contact the service
organization to readjust the tip deflection of the transducer. In this way, the risk
of “buckling” or “U-turning” of the transducer in the esophagus is minimized.
WARNING: Inspect the transducer connector pins for any damage that might
interrupt signal flow through the connection. Be sure to check pins before
connecting transducer to the
ZONARE ultrasound system (see Figure 6-3 on
page 6-3). If pins are recessed, do not use the transducer and call ZONARE
Technical Support.