z.one ultra and z.one ultra sp OPERATOR MANUAL 14-9
c. System displays as
Restore proceeds.
3. Select Exam Type (Figure 14-10) to individually select Presets to be restored
based on transducer and exam type.
Figure 14-14. Restore Presets | Exam Type Screen
a. Select Transducer and exam types in Categories box and click Apply. The
following message displays:
Figure 14-15. Overwrite Prompt
b. Select Ye s . (Selecting No aborts the restore.) The system restores the
Preset. You may select another transducer/exam to restore or click Cancel
to return to imaging.
4. Select Single Preset (Figure 14-10) to select individual Presets to be restored.
NOTE: If desired, uncheck the options you do not want to Restore. If so, the
following message displays when you click Restore.
Figure 14-13. Preset Restore Prompt
Select Ye s to proceed with restore. (Selecting No will abort the restore).