7
7-80 z.one ultra and z.one ultra sp OPERATOR MANUAL
Optimize
SmartCart/SmartCart sp
Overall / DGC Gain
■ Press the Optimize button to automatically balance the Overall/DGC gain. The
B-Mode image will adjust the brightness of the image to the default Target Gain
value.
Sound Speed Correction
■ Press and hold down the Optimize button to automatically compensate for the
sound speed in tissue. The B-Mode image will pause momentarily, then adjust
for the detected sound speed.
To Exit Optimize Mode
■ Double-click the Optimize button to turn Optimize functions = OFF.
Compounding
SmartCart/SmartCart sp
SmartCart
NOTE: All ZONARE ultrasound factory Presets have Compounding as the
default. To adjust B-Mode frequencies, first exit Compounding as follows:
■ Highlight Compounding in the Imaging menu. Press Select to turn OFF the
Compound Harmonic imaging function. Repeat this process to turn ON
Compounding.
■ There may be more than one Compounding Image selection depending on the
transducer. To select another Compounding setting, press the left/right arrows
on the Menu Control.
NOTE: On the SmartCart, press the Frequency button up or down when doing
Compound imaging to cycle through Frequency and Compounding choices.
A Output ■ Press the Menu button. Highlight A Output with the Trackball. Move the
Trackball left/right (or press left/right Set buttons) to decrease or increase the
desired transmit power output level.
■ The current transmit power level is displayed (as a percentage; 25%, 50%, etc.)
in the Image Information area. The default value is 100%.
Tints ■ Press Menu button. Highlight Tints with Trackball. Move Trackball Left / Right
to select desired Tint to be applied to image.
NOTE: Tints may be applied separately to Dual images.
Table 7-8. Contrast Imaging User Controls (Continued)
Control SmartCart & SmartCart sp