10
10-6 z.one ultra and z.one ultra sp OPERATOR MANUAL
The ECG trace is part of the Cardiac Exam Preset and automatically scrolls at the
bottom of the 2D image. In strip modes (M, PW, and CW modes) the ECG trace
will default to the top of the strip. The ECG trace position can be adjusted
anyplace within the strip box.
WARNING: The ECG is not intended for ECG diagnosis. It must not be used for
intraoperative applications of the heart. Use only the recommended patient cable
supplied by ZONARE. Make sure that bare parts of the electrodes and the patient
do not come in contact with conductive parts, such as metal examination beds,
trolleys, and similar items.
WARNING: When using the ECG input and defibrillation is required, always
disconnect the ECG cable from the system before defibrillation and keep cable
away from conductive surfaces.
NOTE: ZONARE recommends the following ECG patient cables and lead wires
from Advanced Medical Cables (www.advantagemed.com):
• AMC LW-3700024/31 (International patient lead wire replacements)
• AMC LW-3700024/3A (Domestic patient lead wire replacements)
• AMC CB-83340 (ECG trunk cable)
• AMC CB-33598-00 (Accessory cable for use without lead wires)
To Use ECG Function
■
A transducer (such as the P4-1c or P10-4 transducer) needs to be initialized
for the ECG to operate.
■
Plug the ECG cable into the ECG port on the SmartCart (see Figure 2-1 on
page 2-2) to display the ECG trace.
■
Connect the three patient leads to the patient.
To Activate ECG Trace
1. Press the ECG key on the SmartCart QWERTY keyboard (Figure 10-4). This
will activate the ECG softkey controls (Figure 10-5).
Figure 10-4. ECG Key on SmartCart QWERTY Keyboard