9-22 z.one ultra and z.one ultra sp OPERATOR MANUAL
Only a single PI measurement can be performed at any given time.
NOTE: In the DDB, Max velocity is identified as PSV (Peak Systolic Velocity) and
Min velocity is identified as MDV (Minimum Diastolic Velocity).
To Measure Pulsatility Index (PI)
1. Freeze the display with a PW Doppler strip image.
2. Press the Measure button.
3. Highlight and select PI.
4. The first PW trace caliper will be displayed.
5. Use the Trackball to position the caliper to the desired location within the PW
Doppler strip display. There is no DDB display at this time.
6. Press the Set button to fix PW trace start point and activate the second
velocity caliper.
7. Use the Trackball to trace the desired portion of the waveform within the PW
Doppler strip display. There is no DDB display at this time.
8. To undo any portions of the trace:
9. Press the Set button to fix the trace end point and complete the PW trace and
activate the PSV velocity caliper. At this time, the DDB only displays the TA
max and the PSV velocity results. The system continuously updates the PSV
velocity result as the user positions the caliper within the strip image.
NOTE: When the trace is completed, a vertical line is dropped from the start and
end points of the trace to the baseline to clearly delineate the traced region used
for TA max.
NOTE: The PSV velocity caliper is automatically positioned at the maximum traced
velocity.
10. Use the Trackball to position the caliper to the desired location within the PW
Doppler strip display.
NOTE: If the absolute value of the MDV velocity result is greater than the current
PSV result, the MDV and PSV result displays are swapped and the caliper
becomes a PSV caliper. The converse is also true when the absolute value of the
PSV velocity result is less than the current MDV result.
NOTE: Pressing the Set button again toggles the active caliper selection between
the first and second velocity calipers, allowing positional corrections to be made to
both caliper instances within the strip image.
SmartCart
Press the Delete button
SmartCart sp
Press the Back/Delete button
Scan Engine
Press the Back button
SmartCart/SmartCart sp
Only updates the PSV velocity result as the user
positions the caliper within the strip image
Vue de la page 223