z.one ultra and z.one ultra sp OPERATOR MANUAL 16-1
16
16
Specifications
This chapter provides electrical, connection, physical, environmental, and
authorized accessory specifications, including the P8-3
TEE transducer.
For other transducer-related specifications, see “Acoustic Output” in the 4.3
Safety Manual.
Component Specifications
Table 16-1. Specifications - SmartCart/SmartCart sp w/Scan Engine
Type Parameter Value
Electrical Power requirements 100 - 240V~, 50 - 60Hz, 6A max
Power consumption (no
peripherals) (max.)
100W (342 BTU/Hr)
Power consumption (w/
peripherals) (max.)
442W (1512 BTU/Hr)
Environmental
(operating)
Cooling requirements See power consumption above.
Air temperature
0 - 35
o
C (32 - 95
o
F)
Humidity 15 - 80%, non-condensing
Pressure 700 - 1060 hPa
Environmental
(storage)
Air temperature
-20 - 60
o
C (-4 - 140
o
F)
Humidity 15-90%, noncondensing
Pressure 500 - 1060 hPa