z.one ultra and z.one ultra sp OPERATOR MANUAL 6-25
WARNING: To avoid injury to the patient, use only the Ocular Preset when
performing imaging through the eye. The FDA has established lower
acoustic energy limits for ophthalmic use. The system will not exceed these
limits only if the Ocular Preset is selected. See the Safety Manual for
information about the lower acoustic energy limits established for
ophthalmic use.
CAUTION: Cardiac rhythm disturbances during perfusion studies using gas
ultrasound contrast agents have been observed in the diagnostic range of MI
values. See the specific Instructions For Use for the contrast agent being used for
details.